Setup of LabNEXT Washing
/ Drying Module
LabNEXT washing / drying unit consists of two chambers –
washing and drying encapsulated in a single body. The washing chamber filled
with washing liquid, usually water or 10% ethanol. In order to provide
circulation of the washing liquid the washing chamber has two fittings used to
connect IN and OUT pipes. The lower pipe used for feeding the chamber with
clean liquid and the higher pipe is for removing the liquid from the chamber.
The height of the OUT pipe limits the level of liquid in the washing
chamber. There are two ways of feeding
the washing chamber with the liquid and removing used liquid from it.
1. Free Flow Feeding from a reservoir that
places higher than the level of liquid in the washing chamber. In this case the
IN tube connects to the reservoir and the OUT tube to vacuum outlet. Vacuum is
used for removing used liquid from the chamber. The flow control valve intended
for adjusting the flow of the liquid from the reservoir and prevent excessive
flow that can cause overflow of the chamber. The flow of the liquid may vary
depend on the height the reservoir is installed. In order to prevent washing
chamber from overflowing the vacuum power must be sufficient to consume amount
of water that comes form the water reservoir. The cut-off valve used to shut-off
feeding of the chamber with the washing liquid without changing settings of the
flow control valve. Shut down the washing liquid first and only after that turn
off the vacuum source.
(When starting operations first turn on the vacuum and then allow the
washing liquid to run into the chamber).
2.
Pump Feeding with a two contour (two head) peristaltic pump.
In this mode one contour is used for feeding the washing chamber with clean
washing liquid from the reservoir. It mast be connected to the lower IN
fitting. The second contour connecter to the OUT fitting used for removing the
liquid from the chamber. In order to prevent the chamber form overflowing the OUT flow must exceed
the IN flow. This can be reached either by using peristaltic heads
with different diameters of the tubes or by using the flow control valve with
the IN tube.
Drying
of the printing needles performed in the drying chamber with vacuum. Regardless
of the installation type the drying chamber must be connected to the vacuum
outlet. It is recommended using a vacuum catcher to prevent liquid particles
from reaching the vacuum system and avoid its contamination.